Chang I. L., Klee M. S. Purge and Trap GC-MS Determination of Trace Volatile Compounds in Drinking Water Using the HP 5973 Series MSD and Volatiles interface
2001 – Zurnal Hromatografichnogo Tovarystva – V. 1. – № 2. – P. 32-55
CHANG I. L., KLEE M. S.
PURGE AND TRAP GC-MS DETERMINATION OF TRACE VOLATILE COMPOUNDS IN DRINKING WATER USING THE HP 5973 SERIES MSD AND VOLATILES INTERFACE- article
Eighty-four volatile pollutants in drinking water were analysed using the HP 6890/5973 GC-S system with P&T and VI. Test parameters were optimised to maximise sensitivity, linearity, and reproducibility. The system was more sensitive, more inert to active compounds, and handled higher column /low (for smaller split ratio and larger volume of sample) than previous systems. The linearity of all 84 analytes was excellent with correlation coefficients of 0.994 to 0.999. The reproducibility of results is comparable to those specified by EPA method 524.2. The RSDs are below 10 percent for most analytes and under 20 percent for active and difficult-to-purge compounds. The method detection limits for most analytes are under 0.05 ppb, which are below the values reported in EPA method 524.2 and the referenced literature.